Analytic deficiencies (D5400s and up)

The analytic phase (42 CFR 493.1250–493.1283) holds most of CMS's top-cited tags: D5413 conditions (818 labs in 2024), D5401 and D5403 procedure manual (635 and 609), D5417 reagents (607), D5421 performance verification (451) and D5429 maintenance (449). D5400 is the analytic systems condition.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

D5400 analytic phase CLIA deficiencies
D5400 analytic phase CLIA deficiencies
TagRequirementTypical gapPrevention / fix
D5401Procedure manual followed, §493.1251(a)Staff practice differs from written procedureAlign procedure with practice, director approval, annual review
D5403Procedure manual content, §493.1251(b)Missing required elements (acceptability, steps, QC, reporting)Content checklist from §493.1251(b) applied to every procedure
D5413Conditions defined, monitored, documented, §493.1252(b)Temperature or humidity not recorded, no rangesDefined ranges, continuous monitoring, weekend coverage
D5417No expired or deteriorated materials, §493.1252(d)Expired controls or reagents in useReceiving log, open-date labels, monthly sweep
D5421Verify performance specs, §493.1253(b)(1)New analyzer used before verificationGo-live hold until director signs verification
D5429Maintenance per manufacturer, §493.1254(a)(1)Maintenance not documentedShift checklist, weekly supervisor review
Calibration§493.1255Six-month calibration verification missingCalendar with event triggers
QC§493.1256Out-of-range QC with results releasedQC lockout or hold rule, documented corrective action, patient look-back

Want a gap check before the next survey?

Sound Lab Consultants reviews laboratories phase by phase against CLIA requirements and helps fix D-tag deficiencies after a survey.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What is the analytic systems condition?

D5400, under 42 CFR 493.1250, requiring the lab to meet analytic requirements in 493.1251–493.1283 and monitor and correct analytic problems.

Why is D5413 cited so often?

It covers every condition essential to test performance and storage, so any missing temperature or humidity record or undefined range can trigger it.

Sources