Most CA 125 citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Where CA 125 citations come from

Survey findings about CA 125 almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.

AreaRegulationRecords surveyors ask for
QC42 CFR 493.1256Daily CA 125 QC records, Levey-Jennings charts, corrective action log, control lot ranges
Calibration verification42 CFR 493.1255Six-month CA 125 calibration verification and reportable range data with director review
Proficiency testing42 CFR 493.801, Subpart HCA 125 PT enrollment, attestation statements, graded reports, investigation of every miss
Method verification42 CFR 493.1253CA 125 verification study: accuracy, precision, reportable range, reference interval check, director sign-off
Test report and critical values42 CFR 493.1291Critical value list, call log with read-back, report format showing CA 125 reference intervals
PersonnelSubpart MTraining and competency records for everyone who runs CA 125

PT status for CA 125

CA 125 (Cancer Antigen 125) is a regulated analyte (endocrinology, 42 CFR 493.933). CLIA PT acceptance limit: target value ±20%.

Specimen problems that turn into CA 125 findings

Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.

If CA 125 was cited on your survey

  1. Identify the exact standard cited and the records the surveyor reviewed.
  2. Look back at patient results that could have been affected and decide whether corrected reports are needed.
  3. Fix the system, not just the instance, and assign monitoring.
  4. Write the plan of correction. See the CLIA plan of correction guide.

Related tests

Full guide: CA 125 (Cancer Antigen 125) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Want a gap check before the next survey?

Sound Lab Consultants reviews laboratories phase by phase against CLIA requirements and helps fix D-tag deficiencies after a survey.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What are the most common CA 125 survey deficiencies?

Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for CA 125. These are general categories, not statistics.

Which regulation covers CA 125 quality control?

42 CFR 493.1256 sets the control procedure requirements.

Is CA 125 on the CLIA PT analyte list?

CA 125 (Cancer Antigen 125) is a regulated analyte (endocrinology, 42 CFR 493.933). CLIA PT acceptance limit: target value ±20%.