Home › Cystatin C

Cystatin C: Survey Deficiencies and CLIA Citations

Most Cystatin C citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Where Cystatin C citations come from

Survey findings about Cystatin C almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.

AreaRegulationRecords surveyors ask for
QC42 CFR 493.1256Daily Cystatin C QC records, Levey-Jennings charts, corrective action log, control lot ranges
Calibration verification42 CFR 493.1255Six-month Cystatin C calibration verification and reportable range data with director review
Accuracy verification42 CFR 493.1236Twice-yearly Cystatin C accuracy checks (external PT, split samples or reference material) with acceptance criteria
Method verification42 CFR 493.1253Cystatin C verification study: accuracy, precision, reportable range, reference interval check, director sign-off
Test report and critical values42 CFR 493.1291Critical value list, call log with read-back, report format showing Cystatin C reference intervals
PersonnelSubpart MTraining and competency records for everyone who runs Cystatin C

PT status for Cystatin C

Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).

Specimen problems that turn into Cystatin C findings

Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.

If Cystatin C was cited on your survey

  1. Identify the exact standard cited and the records the surveyor reviewed.
  2. Look back at patient results that could have been affected and decide whether corrected reports are needed.
  3. Fix the system, not just the instance, and assign monitoring.
  4. Write the plan of correction. See the CLIA plan of correction guide.

Related tests

Full guide: Cystatin C testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Want a gap check before the next survey?

Sound Lab Consultants reviews laboratories phase by phase against CLIA requirements and helps fix D-tag deficiencies after a survey.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What are the most common Cystatin C survey deficiencies?

Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for Cystatin C. These are general categories, not statistics.

Which regulation covers Cystatin C quality control?

42 CFR 493.1256 sets the control procedure requirements.

Is Cystatin C on the CLIA PT analyte list?

Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).