Glucose: Survey Deficiencies and CLIA Citations
Most Glucose citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where Glucose citations come from
Survey findings about Glucose almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily Glucose QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month Glucose calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | Glucose PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | Glucose verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Glucose reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Glucose |
PT status for Glucose
Glucose is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±8% or ±6 mg/dL (greater).
Specimen problems that turn into Glucose findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Glycolysis lowers glucose in unseparated whole blood; separate promptly or use a tube with a glycolysis inhibitor as your procedure specifies.
- Glucose meters cleared for home use are excluded from the regulated analyte; their use in a lab still needs proper certificate coverage and QC.
- Contamination from dextrose IV lines raises results.
If Glucose was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- Hemoglobin A1c (HbA1c) (full guide)
- Potassium
- Sodium (full guide)
- Beta-Hydroxybutyrate (Ketones) (full guide)
Full guide: Glucose testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a gap check before the next survey?
Sound Lab Consultants reviews laboratories phase by phase against CLIA requirements and helps fix D-tag deficiencies after a survey.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common Glucose survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for Glucose. These are general categories, not statistics.
Which regulation covers Glucose quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is Glucose on the CLIA PT analyte list?
Glucose is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±8% or ±6 mg/dL (greater).
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)