Urinalysis (Dipstick and Microscopic): Survey Deficiencies and CLIA Citations
Most Urinalysis citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where Urinalysis citations come from
Survey findings about Urinalysis almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily Urinalysis QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month Urinalysis calibration verification and reportable range data with director review |
| Accuracy verification | 42 CFR 493.1236 | Twice-yearly Urinalysis accuracy checks (external PT, split samples or reference material) with acceptance criteria |
| Method verification | 42 CFR 493.1253 | Urinalysis verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing Urinalysis reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs Urinalysis |
PT status for Urinalysis
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).
Specimen problems that turn into Urinalysis findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Test urine promptly or refrigerate; delayed testing changes cells, casts, bacteria and pH.
- Reagent strips must be within date, properly stored, and read at the timed interval.
- Microscopic urinalysis by a provider may fall under a PPM certificate; dipstick-only testing may be waived.
If Urinalysis was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- Urine Albumin (Microalbumin) and Albumin/Creatinine Ratio (full guide)
- Urine Pregnancy Test (Urine hCG) (full guide)
- Urine Drug Screen (Drugs of Abuse)
Full guide: Urinalysis (Dipstick and Microscopic) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a gap check before the next survey?
Sound Lab Consultants reviews laboratories phase by phase against CLIA requirements and helps fix D-tag deficiencies after a survey.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common Urinalysis survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for Urinalysis. These are general categories, not statistics.
Which regulation covers Urinalysis quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is Urinalysis on the CLIA PT analyte list?
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)).
Sources
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)